H303-T-H-CO2/O2[0-21]

T-H-CO2/O2[0-21] StageTop Controller set contains the temperature controller for BL3 electrically heated chambers, active humidity, CO2 (0-20%) and O2 (0-21%) controller, air pump, the user touch screen interface and the USB cable for LASX connection.

The temperature module allows to work in chamber or in specimen feedback mode and to continuously monitor room temperature. T range: 3°C above ambient – 60°C. Specimen Temperature accuracy: ± 0.1°C in specimen feedback mode, ± 0.3°C in chamber feedback mode, if room temperature remains within ± 1°C. Automatic self-calibration routines for the incubating chamber and for the objective heater.

Oxygen and CO2 concentration are digitally controlled by mixing Air (from the Okolab air pump), N2 and CO2. CO2 range: 0-20%; Oxygen range:0-21%. CO2 sensor: Non Dispersive InfraRed (NDIR) dual wave length detector. Expected lifetime: 10 years. Accuracy: ± 0.1%. Oxygen sensor: Optical Oxygen sensor. Expected lifetime 5 years. Resolution of 0.1%, accuracy of ± 0.1% at 5% Oxygen. Output flow rate range: 0.2-0.4 l/min, with factory setting at 0.4 l/min.

The Active Humidity Controllers employs a humidity sensor to control gas relative humidity. Operating thermal range: 25÷50°C. Humidity range depends on working temperature. At 37°C humidity can be regulated in the range 50 to 95%. Humidity Set-point Resolution: 1%. It features a temperature controlled tube to deliver the humidified gas to the incubating chamber.

Integrated into LASX 2.0.

Compatible with the CO2-H-IN-CHAMBER SENSOR for in chamber closed loop control on CO2, O2 and Humidity.

back